§882.1935. Near Infrared (NIR) Brain Hematoma Detector.

21 C.F.R. § 882.1935

(a)
Identification. A Near Infrared (NIR) Brain Hematoma Detector is a noninvasive device that employs near-infrared spectroscopy that is intended to be used to evaluate suspected brain hematomas.
(b)
Classification. Class II (special controls). The special controls for this device are:
(1)
The sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2)
The labeling must include specific instructions and the clinical training needed for the safe use of this device;
(3)
Appropriate analysis/testing should validate electromagnetic compatibility (EMC), electrical safety, and battery characteristics;
(4)
Performance data should validate accuracy and precision and safety features;
(5)
Any elements of the device that may contact the patient should be demonstrated to be biocompatible; and,
(6)
Appropriate software verification, validation, and hazard analysis should be performed.